Table 3. Alignment tolerance, precision and range of mechanical motion.
Note that we adopt the SHADOW convention for the coordinates: on the optical surface, X is transverse to the beam path, Y is along the beam path, and Z is normal to the optical surface.
| Alignment tolerance | Mechanical motion (step resolution; range) | |
|---|---|---|
| Spherical mirror | ||
| MX (transverse) | ±25 µm | N/A |
| MY (along beam path) | ±25 µm | N/A |
| MZ (up/down) | ±5 µm | N/A |
| MΘX (pitch) | ±0.01° | 0.0001°/step; 2° |
| MΘY (roll) | ±0.01° | N/A |
| MΘZ (yaw) | ±0.1° | N/A |
| VLS plane grating | ||
| GX (transverse) | N/A | 10 µm/step; 40 mm |
| GY (along beam path) | ±50 µm | N/A |
| GZ (up/down) | ±10 µm | N/A |
| GΘX (pitch) | N/A | 0.0001°/step; 5° |
| GΘY (roll) | ±0.01° | N/A |
| GΘZ (yaw) | ±0.1° | N/A |
| Spectrometer | ||
| SX (transverse) | ±100 µm | N/A |
| SY (along the beam) | ±100 µm | N/A |
| SZ (up/down) | ±10 µm | N/A (or source vertical move) |
| Detector | ||
| DY (along the beam) | N/A | 1 µm/step; 150 mm |
| Wolter mirrors WM1 | ||
| WM1X | 100 µm | N/A |
| WM1Y | ±5 µm | 0.5 µm/step; 1 mm |
| WM1Z | ±2 µm | 0.2 µm/step; 100 µm |
| WM1ΘX | ±0.001° | 0.0001°/step; 0.1° |
| WM1ΘY | ±0.01° | N/A |
| WM1ΘZ | ±0.01° | N/A |
| Wolter mirrors WM2 | ||
| WM2X | 100 µm | N/A |
| WM2Y | ±10 µm | 0.5 µm/step; 1 mm |
| WM2Z | ±2 µm | 0.2 µm/step; 100 µm |
| WM2ΘX | ±0.001° | 0.0001°/step; 0.1° |
| WM2ΘY | ±0.005° | N/A |
| WM2ΘZ | ±0.01° | N/A |