Extended data Figure 3. Ruthenium isotope data obtained for ng ml-1 Alfa Aesar Ru standard solutions doped with variable amounts of Ni and S.

Extended data Figure 3

The effect of Ni argide interferences on the measured ε100Ru (a) and ε98Ru (b) isotopic compositions of a 100 ng ml-1 Ru standard solution doped with varying amounts of Ni. The accuracy of the measured isotopic compositions is not affected for samples with Ni/Ru < 10-2. Panels (c) and (d) show that even if high amounts of S (S/Ru = 5) are present in the analyzed sample solutions, this has no effect on the accuracy of the measured ε100Ru and ε96Ru data. Vertical dashed lines in (c) and (d) indicate the range of S/Ru of the analyzed samples. The blue areas in (a-d) indicate the external reproducibility of the method as defined by the standard deviation (2 s.d.) of replicate digestions and repeated analysis of the UG-2 reference sample (see Methods).