35 |
Different volumes in tubes after centrifugation |
Evaporation during ultracentrifugation |
Test all buckets with test gradients without lysate, marking the level of liquid in the tubes. Avoid using buckets for which evaporation is observed. |
|
Variability in number of fractions recovered |
Differences between pipet sets or users |
Try to use always the same pipet set for every replicate. The replicates should also be performed by the same user. |
42 |
Difficulty to detect protein bands |
Less than 1 mg lysate was loaded onto gradient; low sensitivity of antibody; low expression of protein |
Load 1–2 mg lysate onto gradient; try an alternative antibody; test control proteins to verify western blot procedure; check protein expression in lysate. |
49 |
Graph is inverted |
The ImageJ software gave back inverted mean values |
Subtract the inverted mean from 255 (for 8-bit images). |
|
Negative protein amounts |
Excel mixed up decimal separator “comma” and “point” |
Check all mean values again; values cannot be higher than 255 (for 8-bit images). |
88 |
TMT label incorporation is <95% |
Poor labeling efficiency |
Repeat labeling (steps 76–82) for all channels: add another aliquot of TMT reagent to each sample for 1 h and proceed with testing and quenching as described (Steps 83–88). |
95 |
No proteins / peptides left |
Pellet got lost in one of the TCA washing steps |
Tubes should be placed into the centrifuge always with the same orientation so that it is clear where the pellet is located. |
|
Low protein / peptide concentration |
Warm acetone was used for washing steps during TCA precipitation; pellets were overdried and could not be dissolved in the next steps |
Acetone needs to be cold to prevent washing away proteins; decrease time for drying, or incubation in an ultrasonic bath with ice can help to re-dissolve the pellet. |