Figure 5.

Figure 5

Comparison with reference measurements in a volunteer at rest. A, T1, T2, and M0 maps obtained from an interleaved undersampled (R = 20) spiral MRF acquisition. B, Water T1 and M0 maps obtained from a fat‐suppressed IR, water T2 maps obtained from an MSE sequence with a tri‐exponential fit, and water and fat fraction (F) maps obtained from DIXON (all Cartesian). The water T1 map obtained with MRF is close to that obtained with fat‐suppressed IR. The T1 map obtained with IR shows a bright region, for which the fat‐suppression pulse was probably not fully effective. The water T2 maps obtained with the MSE sequence show shorter values compared with those obtained from the MRF measurements. Note that the fat‐suppressed reference measurements do not deliver information about the fat