Table 2.

Complete results for maximum achievable IDR vs injection system vs contrast media vs catheter gauge: values represent the average of three trials for each combination of contrast media, injection system, and catheter gauge

Catheter gauge Contrast media Concentration (mgI/mL) Maximum iodine delivery rate (gI/s)
Centargo Stellant MP CT motion CT Exprès
18G Iopromide 300 3.00 3.00 2.70 1.59
Iodixanol 320 2.27 2.18 2.02 1.18
Iohexol 350 2.91 2.77 2.49 1.40
Iopromide 370 3.15 3.07 2.70 1.63
Iomeprol 400 2.48 2.40 2.14 1.32
20G Iopromide 300 2.70 2.61 2.12 1.38
Iodixanol 320 1.76 1.66 1.54 1.00
Iohexol 350 2.21 2.07 1.89 1.18
Iopromide 370 2.41 2.29 1.92 1.37
Iomeprol 400 2.10 2.02 1.80 1.09
22G Iopromide 300 1.65 1.56 1.26 0.99
Iodixanol 320 1.12 1.02 0.78 0.66
Iohexol 350 1.37 1.16 0.98 0.81
Iopromide 370 1.48 1.41 1.13 0.83
Iomeprol 400 1.28 1.16 0.88 0.74

IDR, iodine delivery rate.