Skip to main content
. 2020 May 8;13(5):100122. doi: 10.1016/j.waojou.2020.100122

Table 1.

Measurements of different skin prick test techniques (Section 1).

Histamine wheal, mean ± SDa Histamine flare, mean ± SDa CV % İntrapatient CV % İnterpatient Sensitivity % (95% CI) Specificity % (95% CI) Pain median (IQR)
Oryum-VP 5.25 ± 0.65a 21.84 ± 4.59a 10.72 27.16 100 (80–100) 100 (80–100) 4 (2–5.5)a
Oryum-VC 7.85 ± 1.19b 33.37 ± 3.6b 14.43 29.65 100 (80–100) 95 (73.1–99.7) 4 (4–6)ab
Oryum-VCC 7.86 ± 1.33b 31.93 ± 4.1b 12.32 29.36 100 (80–100) 95 (73.1–99.7) 5 (4–8)b
Oryum-WP 4.22 ± 0.89c 18.99 ± 4.5c 24.31 37.26 100 (80–100) 100 (80–100) 4 (2–6)ab
Lancet 5.45 ± 0.85a 24.85 ± 5.42a 11.99 28.88 100 (80–100) 100 (80–100) 5 (2.5–8)ab
P <0.001 <0.001 0.008

a, b, c: different letters in the same column represent the statistically significant difference (p < 0.05).

a

Values for wheal and flare expressed in millimeters