The lowest mass of explosives detected in each replicate measurement from various surfaces. Key: N/D = not detected.

Lowest detected mass (ng)
Tsai et al. [33], recovery from glass slide
This work, recovery from glass slide
Clean keyboard
Used keyboard
Paper spray Paper Spray Swab Spray Swab Spray Swab Spray
TNT 800 N/D N/D ND Partial at 25
RDX 100 25 10 10 10
HMX 600 25 10 Partial at 10 10
PETN 100 25 10 Partial at 10 10
Tetryl 25 10 Partial at 10 10
NG N/D N/D ND 25
PA 10 10 10 10