Table 2.
Comparison of SEV exhaled by the patient and SEV in the patient’s expiratory hose after the carbon filter
| ACD-100 | ACD-50 | t | ACD-100 | ACD-50 | t | ANOVA | Bonferroni | |
|---|---|---|---|---|---|---|---|---|
| 30 min | 90 min | p | 60 min | 120 min | p | p | Post hoc | |
| Sevoflurane exhaled by patient (%) | 0.91 ± 0.17 | 0.91 ± 0.21 | 1.000 | 0.99 ± 0.16 | 0.78 ± 0.20 | < 0.001 | 0.003 | 60 × 120 |
| Sevoflurane lost (%) | 0.07 ± 0.07 | 0.19 ± 0.08 | < 0.001 | 0.11 ± 0.08 | 0.16 ± 0.09 | 0.116 | < 0.001 |
30 × 60 30 × 120 60 × 120 |
| SRI | 95.29 ± 6.45 | 85.54 ± 11.15 | 0.001 | 93.42 ± 6.55 | 88.77 ± 12.26 | 0.135 | 0.003 |
30 × 90 60 × 90 |
P values of Student’s T-test for paired samples and ANOVA p values with Bonferroni post hoc correction. Bonferroni shows between which groups of minutes there are differences
SRI Sevoflurane reflection index