Table 2.

Dynamic linear range, recovery, and matrix effects observed for AQ standards at 25 nM (LLOQ), low QC (50 nM), middle QC (250 nM), and high QC (1000 nM) concentrations. Recovery and matrix effects were calculated by comparison of peak areas from pre-extraction spike QC samples (recovery) and post-extraction spike QC samples (matrix effects) with those from QC standards prepared in neat solvent ± coefficient of variation

Dynamic linear range % Recovery (±%CV) % matrix effect (±%CV)
25 nM
(LLOQ)
50 nM
(Low)
250 nM
(Med)
1000 nM
(High)
25 nM
(LLOQ)
50 nM
(Low)
250 nM
(Med)
1000 nM
(High)
C7-PQS 25–1000 nM (R2 = 0.997) 94.3% ± 2.4% 105.8% ± 5.0% 113.4% ± 6.1% 92.0% ± 1.4% 78.7% ± 4.7% 83.1% ± 1.5% 88.2% ± 3.3% 84.7% ± 2.8%
C9-PQS 25–1000 nM (R2 = 0.995) 39.2% ± 4.0% 55.1% ± 2.2% 76.5% ± 5.5% 81.8% ± 2.2% 76.5% ± 3.6% 82.8% ± 3.2% 86.6% ± 3.5% 83.2% ± 4.5%
HHQ 25–1000 nM (R2 = 0.998) 41.3% ± 2.8% 49.3% ± 2.3% 68.2% ± 3.9% 81.9% ± 4.8% 83.3% ± 2.3% 85.1% ± 1.53% 87.1% ± 2.9% 86.5% ± 4.6%
NHQ 25–1000 nM (R2 = 0.998) 68.7% ± 5.7% 80.1% ± 7.0% 80.0% ± 2.1% 83.8% ± 4.9% 129.5% ± 10.5% 129.0% ± 6.4% 123.1% ± 5.6% 122.1% ± 5.0%
HQNO 25–1000 nM (R2 = 0.997) 100.9% ± 4.3% 98.8% ± 3.9% 103.0% ± 5.6% 89.3% ± 5.6% 16.0% ± 1.6% 33.6% ± 1.8% 67.1% ± 2.3% 78.3% ± 3.7%
NQNO 25–1000 nM (R2 = 0.997) 78.0% ± 4.0% 85.6% ± 3.6% 93.7% ± 4.1% 80.2% ± 2.4% 18.3% ± 1.7% 40.7% ± 2.4% 74.5% ± 4.2% 90.2% ± 5.0%