ORs with 95% CIs for the percentage error for the heart rate category and within 10 beats/min for each method, CT and VT, during phases I and II

Phase Accuracy criteria CT (% errors) VT (% errors) OR (95% CI)
I Category 65/220 (29.5) 36/220 (16.3) 2.1 (1.4 to 3.4)
±10 beats/min 75/220 (34.1) 32/220 (14.5) 3.0 (1.9 to 4.8)
II Category 27/135 (20.0) 17/135 (12.6) 1.7 (0.9 to 3.4)
±10 beats/min 27/135 (20.0) 8/135 (5.9) 4.0 (1.7 to 8.5)
Total category 92/355 (26%) 53/355 (15%) 2.0 (1.4 to 2.9)
Total ±10 beats/min 102/355 (29%) 40/355 (11%) 3.2 (2.1 to 4.8)