Skip to main content
. 2020 Jan 9;40(9):1323–1330. doi: 10.1038/s41372-019-0586-4

Table 3.

Validation for the detection of staphylococcal enterotoxins A (SEA), B (SEB) and D (SED) using ELISAc.

Parameters SEA SEB SED
Accuracy, % 7.8 12 10
Precision, %CV 5.4 2.7 7.6
limit of detection, ng/mL 0.25 0.43 0.43
limit of quantification, ng/mL 0.54 2.4 0.41
LLOQa, ng/mL 0.16 0.83 0.16
ULOQb, ng/mL 200 200 200

aLLOQ, lower limit of quantification

bULOQ, upper limit of quantification

cValidation values were from supernantant human milk sample containing 25 ng/ml of SEA, SEB or SED with 4 days and 3 replicated for each day