Table 2:

Calibration equations for predicting systolic and diastolic blood pressure with an oscillometric device using random-zero sphygmomanometer blood pressure measurements as the independent variable.

Approach Calibration Equations
Systolic Blood Pressure Diastolic Blood Pressure
Ignoring the Change OD = RZS OD = RZS
Ordinary Least Squares OD = 9.81 + 0.93*RZS OD = 10.40 + 0.83*RZS
Average Difference OD = 1.09 +RZS OD = −2.44 + RZS
Deming Regression OD = −4.31 + 1.04*RZS OD = −6.07 + 1.05*RZS
Robust Regression OD = 11.02 + 0.92*RZS OD = 10.36 + 0.83*RZS

RZS=random-zero sphygmomanometer, OD=oscillometric device.