Skip to main content
. 2020 May 15;10:8029. doi: 10.1038/s41598-020-64900-w

Figure 1.

Figure 1

Micro-biopsy device design. (a) The micro-biopsy device features a movable cutting tip inside a cutting tube with a depth-limiting flange. The device is inserted into the cardiac ventricle wall until the collar abuts the wall (usually with a penetration depth of 2–3 mm). The tip is then advanced further into the tissue, rotated and retracted to cut and catch a sub-milligram piece of myocardial tissue. (b) Size comparison with clinically used devices. The micro-EMB device does not enlarge during sampling procedure, causing minimal trauma to the cardiac ventricle wall. (C) Ventricular wall surface after ex-vivo application of a conventional EMB device (black arrow) and the micro-EMB device (white arrow) on the left ventricle of swine. Affected regions are outlined below the image, and shows a reduced size of endocardial surface trauma from the micro-EMB device compared to the conventional EMB device (n = 3 replicates).