Table 2.

Bias (SD) of difference, limits of agreement and concordance determined by HemoCue 201+, HemoCue 301 and Sysmex among the pregnant women in North India (N = 102)

Bias (SD) of difference compared to Sysmex XS Limits of agreement Pearson correlation coefficient ρ Lin’s concordance correlation coefficient ρc
HemoCue 201+ (overall) N = 102 − 0.53 (1.01) − 2.52 to 1.45 0.80 0.74
HemoCue 201+ (2nd drop) n = 51 − 0.43 (1.01) − 2.42 to 1.55 0.85 0.80
HemoCue 201+ (3rd drop) n = 51 − 0.63 (1.01) − 2.61 to 1.34 0.73 0.66
HemoCue 301 (overall) N = 102 − 0.25 (0.85) − 1.93 to 1.42 0.85 0.83
HemoCue 301 (2nd drop) n = 51 0.65 (0.72) − 1.35 to 1.48 0.86 0.86
HemoCue 301(3rd drop) n = 51 − 0.57 (0.87) − 2.23 to 1.12 0.88 0.81

SD standard deviation