Skip to main content
. 2020 May 18;22:37. doi: 10.1186/s12968-020-00618-y

Table 1.

bSSFP cine imaging sequence parameters

0.55 T breath-held bSSFP cine 1.5 T breath-held bSSFP cine 0.55 T free breathing re-binned bSSFP cine 1.5 T free breathing re-binned bSSFP cine
Field of view (mm2) 360 × 270 360 × 270 360 × 270 360 × 270
Slice thickness (mm) 8 8 8 8
Matrix size 256 × 192 256 × 140 192 × 108 192 × 119
TE (ms) 1.67 1.2 1.34 1.06
TR (ms) 4.1 2.79 3.24 2.52
Acquired temporal resolution (ms) 32 28 N/A N/A
Bandwidth (Hz/Px) 350 1085 501 1085
Parallel imaging acceleration factor 2 2 3 4
Seconds/slice 9 8 18 16
Calculated Phases 30 30 26 30
Flip angle (°) 78 50 80 50

Sequence parameters for breath-held and free breathing re-binned cine acquisitions at 0.55 T and 1.5 T; bSSFP balanced steady statae free precession, TE echo time, TR repetition time