. 2019 Nov 11;90(Suppl 11):65–71. doi: 10.23750/abm.v90i11-S.8820

Table 1.

Absolute humidity, with and without active humidification, in the investigated steps

Without heated humidification With heated humidication
FiO2 Gas Flow
60 L/min
Gas Flow
80 L/min
Gas Flow
60 L/min
Gas Flow
80 L/min
0.35 Temp° 26.3±0.6 26.5±0.8 29.0±0.0 28.9±0.2
RH (%) 30.8±8.5 § 28.5±5.7§ 54.3±4.4 52.0±3.2
AH (mgH2O/L) 7.7±2.3* 7.1±1.5 15.7±1.4 14.9±0.9
0.5 Temp° 26.0±1.2 26.0±1.3 29.0±0.1 29.0±0.2
RH (%) 26.2±4.9§ 25.0±1.0§ 52.0±3.6 54.5±1.1
AH (mgH2O/L) 6.6±1.6* 5.9±0.2 14.9±1.0 14.1±0.9
0.7 Temp° 25.5±0.9 25.8±1.3 28.9±0.2 29.0±0.0
RH (%) 20.1±2.4§ 21.0±2.8§ 51.0±3.1 51.3±1.5
AH (mgH2O/L) 4.9±0.8* 5.0±0.9 14.6±0.1 14.7±0.4
1 Temp° 25.4±0.7 25.5±1.0 26.0±0.1 29.0±0.0
RH (%) 14.8±2.3§ 13.0±2.2§ 54.2±3.0 49.0±5.2
AH (mgH2O/L) 3.6±0.6* 3.1±0.5 13.4±1.0 14.1±1.5

Legend:

Temp°: temperature, RH : relative Humidity, AH : Absolute Humidity

*repeated measures analysis of variance (rm ANOVA) = p < 0.001

§ repeated measures analysis of variance (rm ANOVA) = p < 0.001