Skip to main content
. Author manuscript; available in PMC: 2020 May 20.
Published in final edited form as: Nat Protoc. 2019 Jul 3;14(8):2483–2520. doi: 10.1038/s41596-019-0188-9
Step Problem Possible reason Possible solution
8 Gradient shim fails at high concentration of Gd-DOTA Gd-DOTA generates a strong PRE effect on the D2O signal Adjust the field to center the deuterium signal, lock and load the shim map used for the previous point of the titration. Perform the gradient shim. If it still fails, shim manually on Z, X, Y, XZ, YZ, and Z.
9 The 20 mM 16-DSA solution turns unclear after a few hours The [16-DSA] is close to its maximum solubility in bicelles Sonicate the solution for 5 minutes right before use; prepare a new 16-DSA solution at lower concentration (~15 mM)