| Step | Problem | Possible reason | Possible solution |
|---|---|---|---|
| 8 | Gradient shim fails at high concentration of Gd-DOTA | Gd-DOTA generates a strong PRE effect on the D2O signal | Adjust the field to center the deuterium signal, lock and load the shim map used for the previous point of the titration. Perform the gradient shim. If it still fails, shim manually on Z, X, Y, XZ, YZ, and Z. |