Skip to main content
. Author manuscript; available in PMC: 2020 May 21.
Published in final edited form as: Nat Protoc. 2018 Sep;13(9):2102–2119. doi: 10.1038/s41596-018-0036-3

Table 1 |.

Troubleshooting table

Step Problem Possible reason Solution
13A(vii) Spheroids do not develop into HOs Insufficient number of medium changes Change the medium every 3-4 d
Absence of growth factors/low growth factor activity Ensure that the growth factors are freshly added
Spheroids sink to the bottom of the dish and fail to develop into HOs The hydrogel is degrading too quickly or is not properly polymerized Reduce the spheroid density per hydrogel Increase the frequency of passages to every 5-7 d
Spheroids were too close to the bottom of the dish during hydrogel cross-linking During hydrogel cross-linking, flip the plate upside-down to ensure that spheroids are not too close to the bottom of the dish
11, 13B(xi) Hydrogels do not form Inadequate pH adjustment of the peptide solutions Adjust the pH to 7.4
The purity level of the peptides was not considered when calculating mass Consider the peptide purity provided by the vendor as shown in Box 1
Determine the peptide purity by UV absorption spectroscopy, as previously described29
Incomplete tissue penetration of the needle bevel at the wound site (only for Step 13B(xi)) Penetrate the tissue slightly past the needle bevel to ensure full delivery of hydrogel solution
13B(xii) Low or no HO engraftment into host tissue Settling of HOs to one end of the syringe due to gravity Shake the syringe gently before each injection
Clogging in the tubing caused by hydrogel cross-linking Untimely mixing of the hydrogel precursor solutions in the tube before injection Verify that there is no leakage from the syringe into the tube before injections
The pH is too high Reduce the pH of the cross-linker solution to delay the reaction during injection