Table 1. LC Gradient Program for Chromatographic Separation of 8-Isoprostanea.

time module event % mobile phase B by volume
0.01 controller solenoid valve  
0.02 controller start  
0.06 autosampler inject  
0.30 pumps %B 40
0.31 pumps %B 55
7.00 pumps %B 65
7.01 pumps %B 100
9.00 pumps %B 100
9.01 pumps %B 40
11.00 controller stop  
a

Mobile phase A is 0.15% formic acid in water and mobile phase B is acetonitrile/0.15% formic acid in water (v/v, 1:1)