Table 5. Underestimation rates of the biopsy devices used.
B-category/Needle | B3 (n = 72) | Underestimation rate | B5a(n = 101) | Underestimation rate | Total underestimation |
---|---|---|---|---|---|
8G | 10 | 2 (20%) | 28 | 3(10.7%) | 5/38 (13.1%) |
10G | 10 | 3 (30%) | 49 | 8 (16.3%) | 11/59 (18.6%) |
14G | 52 | 7(13.4%) | 24 | 9 (37.5%) | 16/76 (21%) |