Table 3:
Troubleshooting
| Step | Problem | Possible Reason | Solution |
|---|---|---|---|
| 1A(ii) | mESCs are not reaching confluency at the predicted time | - Need to seed a different ratio at passage | - The passage ratio can be changed in order to increase cell density or shift when passaging needs to be conducted. Passage at a 1:3 ratio to achieve confluence after 1 day. Passage at a 1:8 ratio to achieve confluence after 3 days. |
| 1B(ii) | hPSCs not reaching confluency | - Initial seeding density is too low | - Try a few different seeding densities when working with a new cell line. |
| 2B(vii) | Cell layers are detaching from plate during human differentiation | - Media is too cold - Re-feeding is dislodging the layers - Incompatible Matrigel lot - Inefficient Matrigel coating |
- Make sure media is at least at room temperature before feeding. - Make sure the pipette dispenser is set to the slowest setting before feeding. - Test the differentiation with different Matrigel lots. - Use fresh Matrigel and be sure to coat for 24 hours. |
| 3A(i) | Not enough V2a interneurons after selection for subsequent experiments | - Need to induce more V2a interneurons | - To obtain the necessary high seeding density, multiple differentiation plates may need to be pooled together depending on the surface for replating or aggregation. |
| 4B(iii) | Cell layers ball up during maturation | The cells have been on the same plate for too long and may be losing the matrix keeping the cells attached to the plate | -Dissociate and replate the cultures onto fresh well plates. Note that this will dilute out the CHX10+ cells. |
| 5A(ix) | Low CHX10 percentage in human differentiations | - Initial seeding density is inappropriate - Incompatible Matrigel lot |
- May need to test different lots of Matrigel to see what supports a neuronal differentiation. - Test a range of seeding densities for each cell line. |
| 5B(i) | Cell layer peeling during staining | Cell layers are releasing during fixation and wash steps | Leave 500 μL of media in the wells and add 500 μL of 4% PFA into the wells to fix. Fix for 1 hour instead of 30 minutes. During wash steps, a transfer pipette may be used to carefully exchange liquids. |
| 5B(vii) | Speckled signal in immunostaining | Excess secondary antibodies aggregate and do not wash out | Filter the solution with the secondary antibody to remove any aggregates. If the target has low abundance, try a lower concentration of secondary antibody (< 1:200). |
| Box 2: 6 | Low cell survival post-thaw | Cell death during thaw | Do not let the vial of cells completely thaw in the water bath. |
| Box 2: 8 | Low cell survival post-thaw | Excess cell debris after thaw | 24 hours after thaw carefully aspirate the media and wash with 0.5 ml PBS. Aspirate PBS and add fresh NIM. |