Table 2.

Clinical sensitivity of the VIASURE qRT-PCR Kit for detection of SARS-CoV-2 as determined in comparative analysis in the University of Birmingham and Queen Elizabeth Hospital. Two commercial assays were used for comparison, the Altona qRT-PCR assay (n = 94) and the Abbott m2000 qRT-PCR assay (n = 26).

Queen Elizabeth Hospital RT-PCRs
Altona Assay
Abbott Assay
Positive Negative Total Positive Negative Total
University of Birmingham VIASURE qRT-PCR Positive 48 1 49 10 0 10
Negative 0 42 42 0 16 16
Indeterminate 0 3 3 0 0 0
Totals 48 46 94 10 16 26