Table 2.

Specifications of the developed human total body PET systems in the Explorer project

System Penn PET Explorer uExplorer
Scintillator LYSO LYSO
Readout SiPM (digital) SiPM
Scintillator size (mm3) 3.86 × 3.86 × 19 2.76 × 2.76 × 18.1
Total number of detector pixels 564,480
Bore/detector diameter 70/81 68.6/78.6
Axial length (cm) 70/140 194.8
Resolution (FWHM in mm)
Transaxial 4.0 3.0
Axial 4.0 3.0–3.5
Energy res. (%) 10 11.7
Energy window 440-660 430-645
Scatter fraction (%)
NU2-2012 32 35.8
Sensitivity
kcps/MBq NU-2 55 191.5 (@0 cm)
Coincidence window (ns) 5 ns 4.5-6.9 ns (ring difference dependent)
TOF resolution (in ps) 250 ps 505 ps
Peak NEC (kcps)
NU-2 -2012 (70-cm phantom) >1200 kcps (for incomplete 70-cm system) 1435 kcps@16.8 kBq/cc
(175-cm phantom) 1718 kcps@8.0 kBq/cc