Table 3.

Diagnostic performance of each B-mode US features and elasticity parameters with optimal cutoff value on SWE for predicting malignancy

Variable Cutoff Sensitivity (%) Specificity (%) PLR NLR AUC
Composition Solid 100 46.30 1.86 0 0.731
Echogenicity Hypoechogenic 82.93 81.48 4.48 0.21 0.838
Margin Irregular 80.49 79.63 3.95 0.25 0.801
Shape Taller than wide 58.54 81.48 3.16 0.51 0.700
Calcification Presence 73.17 85.19 4.94 0.31 0.808
US category High suspicion 95.12 85.19 6.42 0.06 0.913
Emean >54.9 36.59 87.04 2.82 0.73 0.631
Emax >47.1 53.66 70.37 1.81 0.66 0.633
Emin >40.4 31.71 88.89 2.85 0.77 0.613
Eratio >1.94 92.68 29.63 1.32 0.25 0.609

US, ultrasound; SWE, shear wave elastography; PLR, positive likelihood ratio; NLR, negative likelihood ratio; AUC, area under the receiver operating characteristic curve; Emean, mean value for a 2-mm diameter region of interest (ROI) placed in the stiffest region; Emax, maximum pixel value for a ROI; Emin, minimum pixel value for a ROI; Eratio, the ratios of mean stiffness for the lesion-to-normal parenchyma.