Table 2.

Performance comparison of vital sign detection algorithms.

Vital Sign Assessed Vital Sign Algorithms Research Articles Experimental Setup
Range/Subjects/Reference Measurement Method
Results
HR, RR Fast Fourier transform (FFT)
[47,52,60] [47] 0.5–1 m/16/ECG
[60]/9 human
/fingertip pulse oximeter
[52] 15 feet/7 humans/electronic heart-monitor
[47] Error rate: 5%
[60] 1%–5%
[52] Mean error for obstructed RR: 0.3 bpm
Mean error for obstructed RR: 0.14 bpm
Mean error for unobstructed HR: 8 bpm
HR, RR Ensemble empirical mode decomposition (EEMD) and Continuous wavelet transforms (CWT) [12] 0.2–5m/3 human/ECG SNR improvement for RR: 7.5 dB
SNR improvement for HR: 4.8 dB
RR Wavelet transform [8,62] [8] 6 –14 m/2 males, 2 females, 1 actuator/FFT method measurements
[62] 1–2.5 m/-/manual measurement
[8] SNR: (−4.91dB–(−8.28 dB)
Deviation: 0.66%–0.24%
[62] Deviation: 0.19 per minute
HR, RR MTI and Chirp Z-transform (CZT) [4] 1 m/3 human/ECG Error rate BR/HR: 1%–2.4%
RR Multiple Higher Order Cumulant (MHOC) [72] 2–7 m/1 human/manual SNR improvement: 13.8 dB
HR, RR HAPA (harmonic path) [70] 5–15 cm/1 male/pulse-oximeter Error rate (MSE) for HR: 1.83%
HR, RR Spectrum-Averaged Harmonic Path (SHAPA) [71] 5–15 cm/8 human/pulse-oximeter Error rate: 16% improvement over [70]
RR IIR filter [18] 1 m/--/respiration monitor belt (RPM) Correlation co-efficient: 0.909
Absolute error: 0.5 mHz
HR Time Domain Processing Algorithm [16] 0.25–1.25 m/5 human/ECG Error rate: 1.01%–4.32%
HR Time series analysis [15] 1m/apparatus/--
Error rate: 1.26%
HR, RR Pulse-Doppler signal processing technique [19] 5–15 cm/13 human/Polysomnography (PSG) Deviation: 5%
HR Maximum likelihood period estimation [63] 0.3 m/1 human/wearable sensor
Mean Square Error: < −8dB
RR, HR Ensemble empirical mode decomposition (EEMD) [57,61] [57] 3 m–16 m/5 humans/--
[61] 50 cm/--/human tissue model
[57] Error rate: 1.5–3.75%
[61] Error rate: 1.12%
RR, HR Harmonic Multiple Loop Detection (HMLD) [77] 0.64 m/5 male, 5 female/pulse oximeter (HR), manual chest wall count (RR) Error rate (RR): 4.95%
Error rate (HR): 5.06%