. 2019 May 31;46(7):3025–3033. doi: 10.1002/mp.13576

Table 2.

Comparison of PET/CT system performance.

Parameter 5‐ring DMI (this work) 4‐ring DMI7, 8 3‐ring DMI11 Discovery IQ12, 13 Signa PET/MR14, 15
System specifications
Axial/Transaxial FOV (cm) 25/70 20/70 15/70 25/70 25/60
Detector (type) SiPM SiPM SiPM PMT SiPM
Scintillator type LYSO LYSO LYSO BGO LYSO
Scintillator depth (mm) 25 25 25 30 25
Scintillator size (mm x mm) 3.95 x 5.3 3.95 x 5.3 3.95 x 5.3 6.3 x 6.3 3.95 x 5.3
Spatial resolution, FBP R/T/A R/T/A R/T/A Rb/Tb/ Ab R/T/A
1 cm, FWHM (mm) 4.3/4.3/5.0 4.1/4.2 /4.5a 4.4/4.8a /4.8a 4.4c/4.1c/5.3c
10 cm, FWHM (mm) 5.5/4.6/6.5 5.5/4.5/6.0 5.6/5.2a /4.9a 5.8c/4.4c/6.7c
20 cm, FWHM (mm) 7.4/5.0/6.6 7.5/4.9/6.1 8.3a /5.6a /4.7a 8.4c/5.2c/7.3c
Sensitivity (cps/kBq)
Center 20.8 13.2a 7.5c 23.3 23.1
10 cm 20.6 13.1a 20.0 22.5
Counting rate statistics
PNECR (kcps) 266 190a 100c 190a 217a
PNECR activity (kBq/ml) 20.8 22.2 20.6c 9.5a 17.8a
SF at PNECR (%) 40.2 41.4 37a 43.8a
Max. error at PNECR (%) 3.61 3.28 3.65
Image Quality
10 mm (CR/BV; %) 46/9.3 53/10.1 28/6.5 43/4.9a
13 mm (CR/BV; %) 54/7.1 63/7.7 47/5.7 49/4.0a
17 mm (CR/BV; %) 66/5.4 70/5.8 62/5.0 60/3.2a
22 mm (CR/BV; %) 71/4.4 78/4.4 68/4.3 77/2.7
28 mm (CR/BV; %) 85/3.8 87/3.7 66a /3.6 79a /2.2
37 mm (CR/BV; %) 89/3.5 90/3.0 72a /3.1 87/1.9
Average lung error (%) 5.9 6.1 22.2a 3.2

DMI, Discovery MI; SiPM, silicon photomultiplier; PMT, photomultiplier tube; LYSO, lutetium yttrium oxyorthosilicate; BGO, bismuth germanium oxide; FBP, filtered back projection; R/T/A, radial//tangential/axial; FWHM, full width at half maximum; PNECR, peak noise‐equivalent counting rate; SF, scatter fraction; max., maximum; CR; contrast recovery, BV, background variability.

a

Significantly different from 5‐ring DMI value.

b

FBP data are not available so results are for iterative reconstruction.

c

Insufficient data are available to determine significance.