Figure 4:

Figure 4:

Following the PR Logic™ algorithm (Medtronic, Minneapolis, MN, USA), the atrial rate and ventricular rate are compared. Here, due to the fact that the atrial rate exceeds the ventricular rate and as there appears to be some degree of AV dissociation (red box) and also the morphology matches the template of sinus rhythm (blue box), therapy is appropriately withheld during AF.