Skip to main content
. 2020 May 27;129:104476. doi: 10.1016/j.jcv.2020.104476

Table 2.

Imprecision of the LightMix® Modular SARS and Wuhan CoV E-gene assay using the SARS-CoV-2 culture isolate extracts.

Intra-assay Inter-assay
Virus titer (TCID50/mL) Number of positive replicates Mean Cp ± SD
(% coefficient of variation)
Mean Cp ± SD
(% coefficient of variation)
1.8 × 101 3 30.00 ± 0.08 (0.26) 30.18 ± 0.22 (0.72)
1.8 × 10° 3 33.50 ± 0.22 (0.67) 33.64 ± 0.25 (0.75)
1.8 × 10−1 3 36.39 ± 0.32 (0.88) 36.85 ± 0.57 (1.54)