Table 2.

Acceptable dose volume variations and unacceptable deviations among institutions requiring one benchmark resubmission

Criterion failure mode 3DCRT (n = 30) IMRT (n = 15) SIB (n = 17)

≥1 unacceptable deviations 9 (30.0%) 2 (13.3%) 5 (29.4%)
≥2 acceptable variations 17 (56.7%) 8 (53.3%) 7 (41.2%)
Both ≥ 1 and ≥ 2 4 (13.3%) 5 (33.3%) 3 (17.6%)
Isolated hot spot 1 (5.9%)
Did not pass visual review 1 (5.9%)

Abbreviations: 3DCRT = 3-dimensionalconformalradiationtherapy; IMRT = intensity-modulated radiation therapy; SIB = simultaneous integrated boost.