Table 4. Stability of PQ. Data represent mean (± SD).

Conditions Untreated*** treated % remained n
Plasma, glass container, 4°C, 16 hr
30 ng/mL 29.8±2.3 30.2±1.1 101 3
800 ng/mL 796±4 832±14 105 3
Autosampler plastic vials, 21–25°C, 7 days
30 ng/mL 32.8±1.7 29.4±1.1 89.6 4
800 ng/mL 865±27 857±19 99.1 4
Five freeze-thaw cycles
30 ng/mL 31.7±1.3 30.9±1.2 97.5 3
800 ng/mL 844±12 843±12 99.9 3
Plasma, -70°C, 21 months*
3 ng/mL 2.76±0.15 2.78±0.04 101 3
200 ng/mL 200±6 209±5 104 3
Stock, -70°C, 40 months**
  8170±605 8091±338 99 3

* Measured with our previously published method

**measured with UPLC-PDA with peek tubing injection needle.

*** untreated samples were prepared fresh in plastic vials.