Skip to main content
. 2020 May 29;10:8772. doi: 10.1038/s41598-020-65225-4

Table 1.

The energy resolution and resolving power at 25.5 keV.

bending radius diffraction order intrinsic bandwidth source size detector PSF energy resolution resolving power
~11 m Si(311) 0.9 eV 3.2 eV 1.1 eV ~3.8 eV ~6711
~11 m Si(111) 3.5 eV 6.5 eV 2.2 eV ~9.0 eV ~2833
~21 m Si(311) 0.9 eV 6.5 eV 2.2 eV ~7.1 eV ~3591

The presented values are the FWHM values for the corresponding functions.