Skip to main content
. 2020 May 30;509:1–7. doi: 10.1016/j.cca.2020.05.050

Table 3.

Agreement and Cohen’s kappa of the 5 analytical systems under evaluation, using the best cut-off from Youden index for CLIA methods and manufactures’ defined cut-off for Euroimmun and Wantai AbT. Seventy-nine samples were used for the comparison, only 63 samples for the Wantai AbT assay.

IgG Ab results Performance
Maglumi IgG Liaison IgG iFlash IgG Euroimmun IgG
Liaison IgG Agreement = 93.67%Cohen’s kappa = 0.863SE = 0.112
iFlash IgG Agreement = 94.0%Cohen’s kappa = 0.891SE = 0.112 Agreement = 91.1%Cohen’s kappa = 0.794SE = 0.111
Euroimmun IgG Agreement = 96.20%Cohen’s kappa = 0.919SE = 0.113 Agreement = 97.5%Cohen’s kappa = 0.945SE = 0.112 Agreement = 91.1%Cohen’s kappa = 0.799SE = 0.110
Wantai AbT* Agreement = 93.65%Cohen’s kappa = 0.849SE = 0.125 Agreement = 98.41%Cohen’s kappa = 0.961SE = 0.126 Agreement = 92.1%Cohen’s kappa = 0.7879SE = 0.123 Agreement = 98.41%Cohen’s kappa = 0.962SE = 0.126
*

AbT = Total antibodies; SE = standard error.