Table 4.

Recommended start settings for AID mode (AnyDANA-loop)

AnyDANA-Loop Settings Setting
Treatments Safety
Max allowed bolus

Set as the largest typical meal bolus for the participant or per investigator discretion.

Note: If different to the pump’s max bolus, the lowest of the two will apply.

Max allowed carbs Based on largest carb intake during run-in or discussion with participant.
Loop
APS mode Open Loop (Stage 1).
Open APS 0.70
Max U/hr a temp basal can be set to This acts as a safety parameter along with the safety multipliers. Can be set to 5x max daily basal to reduce likelihood of needing to increase it as safety multipliers are increased.
Max IOB Set as the highest bolus from the previous two weeks + 2x their maximum daily basal or lower as per investigator and participant discretion.
Enable SMB Set as ‘off’ by default at the start.
Advanced Settings
Max daily basal safety multiplier

Set as 3x the subject’s maximum daily basal rate (units/hour) or per investigator discretion.

Note: If different to the pump’s max allowed basal, they will operate independently as limits.

Current basal safety multiplier Set at 4x the current basal from the participant’s run-in period or RCT phase if the subject is switching from SAPT to AID for the continuation phase.
Absorption Settings
Minimum 5 min carbohydrate impact Set to 8.
Max autosens ratio Set to 1.2.
Min autosens ratio Set to 0.8.
Other - Temp Targets
Activity duration Set to 120 min.
Activity target Set to target of 8.0.
Eating soon duration Set to a duration of 60 min.
Eating soon target Set to target of 4.5 mmol/L.
Hypo duration The default setting is 30 min.
Hypo target The default setting is 6.5 mmol/L.
Other - Alerts
Pump is unreachable Set to on (default time is 30 min).