Skip to main content
. Author manuscript; available in PMC: 2021 Jan 24.
Published in final edited form as: Phys Med Biol. 2020 Jan 24;65(3):035002. doi: 10.1088/1361-6560/ab6011

Table I.

Average measured contrast over all axial positions

a. Cylindrical phantom
6 mm Spheres 10 mm Spheres
4:1 uptake 6:1 uptake 2:1 uptake 3:1 uptake
Simulation 1.59 ± 0.10 2.02 ± 0.12 1.43 ± 0.06 1.88 ± 0.09
Measurement 1.61 ± 0.15 2.03 ± 0.17 1.43 ± 0.09 1.89 ± 0.11
b. CTN phantom and human subject
6 mm Spheres 10 mm Spheres
Liver (4:1) Lung (8:1) Liver (3:1) Lung (6:1)
CTN Phantom 1.62 ± 0.18 2.45 ± 0.53 1.92 ± 0.12 2.97 ± 0.60
Human Subject 1.54 ± 0.15 2.15 ± 0.34 1.92 ± 0.20 2.84 ± 0.54