Table 8. Comparison of target localization coordinates for an electrode tip in a single case. Here, MidACPC coordinates are presented. All Euclidean distances are compared to the "gold standard" postoperative CT, including the final intraoperative image (O-arm2) and the originally planned target. RT and PP fair well relative to the post-op CT with Euclidean distances of 0.48 mm and 0.30 mm as compared to O-arm2 of 0.41 mm.

AP = Antero-Posterior; LAT = Lateral; VERT = Vertical; MidACPC = Middle of the Anterior Commissure to Posterior Commissure point; RT = Ray Tracing; PP = Perspective Projection; Rel Post-op CT = Euclidian Distance relative to the post-op CT; Rel to O-arm2 = Euclidian Distance relative to the intraoperative O-arm2; Rel to Plan Target = Euclidian Distance relative to original planned target.

  Planned Target O-arm2 CT RT PP
AP -3.5 -3.02 -3.16 -3.32 -3.43
LAT -12 -11.15 -10.8 -10.6 -10.74
VERT -4 -3.71 -3.88 -3.47 -3.76
Rel Post-op CT 1.252996 0.413521 0 0.483425 0.301496
Rel to O-arm2 1.018332 0 0.413521 0.670895 0.581979
Rel to Plan Target 0 1.018332 1.252996 1.507747 1.284562