Table 4.
Results of Bland-Altman Analyses, Performed Between Each MR Modality and Doppler Ultrasound for Each of Three Indices Considereda
| Measured Parameter | Two Modalities | Number of Subjects | Bias | Agreement Limits |
|---|---|---|---|---|
| Peak velocity | DUS and C4DF | 4 | 10.25 | (−36.83, 57.33) |
| DUS and 4DSF | 4 | 2.75 | (−31.85, 37.35) | |
| Time to peak | DUS and C4DF | 4 | −5.75 | (−65.31, 53.81) |
| DUS and 4DSF | 4 | −6 | (−65.47, 53.47) | |
| AV Eject time | DUS and C4DF | 4 | −16.25 | (−53.24, 20.74) |
| DUS and 4DSF | 4 | −20 | (−76.01, 36.01) |
AV, aortic valve; 4DSF = four-dimensional spiral flow, C4DF = conventional 4D flow; DUS = Doppler ultrasound.
Peak velocity; time to peak; AV eject time.