Comparison of patients based on the BP categorization by two different instruments

Automated BP device Mercury-free LED BP Device Validity measures

Hypertensive Non Hypertensive Total
Hypertensive 57 19 76 Sensitivity=96.61%.
Specificity=92.21%,
PPV=75%,
NPV=99%,
Accuracy=93%,
Kappa statistics=0.80
Nonhypertensive 2 222 224
Total 59 241 300