. 2020 Jun 2;70(697):e548–e554. doi: 10.3399/bjgp20X710381

Table 3.

Mean absolute error compared with reference device and failure rate of monitors (n = 327) by frequency of usea

Frequency of use Monitors, n Failing monitors, n (%)b MAE, mmHg (95% CI)
>10 times a month 27 3 (11) 0.89 (0.84 to 0.94)
6–10 times a month 41 2 (5) 0.74 (0.69 to 0.78)
1–5 times a month 115 12 (10) 0.86 (0.84 to 0.89)
6–11 times a year 57 13 (23) 1.08 (1.04 to 1.11)
1–5 times a year 44 10 (23) 0.96 (0.92 to 1.00)
Less frequently 28 5 (18) 0.94 (0.88 to 0.99)
Not stated 15 1 (7) 0.67 (0.60 to 0.74)
a

Four monitors could not be tested fully but should be counted as failures.

b

Percentage of total monitors per frequency-of-use classification. CI = confidence interval. MAE = mean absolute error.