Table 2.

Instrument and working conditions for high-performance liquid chromatography (HPLC) of quercetin, kaempherol, silibinin, and genistein.

Instrument Condition A Condition B
Column CAPCELL PAK C18 MG, 4.6 mm I.D. × 250 mm
Detector UV-Vis Detector
Wavelength 254 nm 350 nm
Analyzed period 30 min 40 min
Solvent Solvent A 2% acetic acid in water Solvent A 0.1% formin acid in MeOH:water = 10:90
Solvent B 0.5% acetic acid in water:ACN = 50:50 Solvent B 0.1% formin acid in MeOH:water = 90:10
Flow rate 1 mL/min 0.4 mL/min
Volume 10 µL 10 µL
Gradient Time (min) Composition (%) Time (min) Composition (%)
A B A B
0 28 72 0 40 60
20 0 100 20 40 60
30 0 100 25 70 30
- - - 40 70 30

Condition A: quercetin, kaempherol. Condition B: silibinin, genistein.