Table 2.

Performance characteristics of the rapid antigen detection test for the presence of SARS-CoV-2 virus in 160 respiratory samplesa.

Sample
type
All samples (N = 160)
High viral load samples (N = 64)
Normal viral load samples (N = 96)
Ct value
No. of samples
Ct value
No. of samples
Ct value
No. of samples
mean range tested positive sensitivity mean range tested positive sensitivity mean range tested positive sensitivity
NPA & TS 20.72 13.80−29.95 35 12 34.3% 16.08 13.80−17.95 11 9 81.8 % 22.84 18.69−29.95 24 3 12.5%
NPS & TS 19.35 13.67−29.79 35 16 45.7 % 16.36 13.67−18.52 20 16 80.0% 23.33 18.79−29.79 15 0 0.0%
sputum 22.91 15.32−31.12 45 5 11.1 % 16.98 15.32−18.14 7 2 28.6 % 24.00 19.43−31.12 38 3 7.9%
throat saliva 19.65 11.25−30.81 45 18 40.0% 15.16 11.25−18.49 26 14 53.8% 25.79 19.20−30.81 19 4 21.1 %
a

High viral load samples means samples with Ct values <18.57 of SARS-CoV-2 virus–specific RT-PCR while normal viral load samples mean samples with Ct values >18.57.