Table 2.

Operative condition of TD-GC/MS analysis.

Step Parameters Value
Tube desorption Purge time 3 min at 5 mL/min–trap in line
Desorption time 10 min
Desorption temperature 300 °C
Temperature of cold trap 20 °C
Desorption flow 30 mL/min, no split
Focusing trap desorption Temperature of cold trap desorption 300 °C
Split low 5 mL/min
Transfer Line Temperature 200 °C
GC analysis Gas carrier He
Gas flow 1.7 mL/min
Analytical column VOCOL® (Supelco), diphenyl dimethyl polysiloxane with crosslinking moieties, 60 m × 0.25 mm ID, 1.5 μm stationary phase thickness
Oven temperature 37 °C hold for 5 min
37–190 °C at 6 °C/min
190–200 °C at 2 °C/min
220–220 °C at 15 °C/min
220 °C hold for 3 min