Skip to main content
. 2020 Mar 20;21(5):48–55. doi: 10.1002/acm2.12852

Table 3.

Patient and tumor characteristics affecting mean quantitative interfraction shifts by setup method.

Outcome Factor Interaction driver and effect on deltas
Vertical Shift BMI

P = 0.034

BMI impacts kV/kV method more than SGRT; increasing BMI leads to increasing vertical shift detected with kV/kV method

AGE

P = 0.035

Age impacts kV/kV method more than SGRT; increasing age leads to increasing vertical shift detected with kV/kV method

Roll Location

P = 0.065

Having a nonlung target location impacts SGRT more than kV/kV method.

BMI

P = 0.022

BMI impacts kV/kV method more than SGRT; increasing BMI leads to decreasing roll detected with kV/kV method

Rotation ROM

P = 0.057

4D ROM impacts kV/kV method more than SGRT; increasing ROM leads to increasing rotation detected with kV/kV method

BMI, body mass index; SGRT, Surface‐guided radiation therapy.