Skip to main content
. 2020 Jun 10;129:104501. doi: 10.1016/j.jcv.2020.104501

Table 1.

Analytical Sensitivity Comparison.

copies/mL* TMA** reactivity Taqpath RT-PCR reactivity CDC RT-PCR††
5.5 × 10e5 5/5 (100 %) 5/5 (100 %) 5/5 (100 %)
5.5 × 10e4 5/5 (100 %) 5/5*** (100%) 2/5 (40 %)
5.5 × 10e3 5/5 (100 %) 0/5 (0%) 0/5 (0%)
5.5 × 10e2 1/5 (20 %) 0/5 (0%) 0/5 (0%)
5.5 × 10e1 0/5 (0%) 0/5 (0%) 0/5 (0%)
5.5 × 10e0 0/5 (0%) 0/5 (0%) 0/5 (0%)
*

Concentration of SARS-CoV-2 genomic RNA, BEI Resources, NR-52285.

**

Hologic Panther SARS-CoV-2 Assay.

ThermoFisher Taqpath COVID-19 Multiplex Assay.

††

CDC 2019-nCoV Real-Time RT-PCR Diagnostic Panel.

***

Specimens were reactive for only one of three genes in assay.