Type II Model Cell parameters (for the components shown in figure 3). (Row 1) Hardware component labels (Voff is zero as there is no battery/voltage-offset component). (Row 2) Patchmaster amplifier software estimates using a simple test pulse. (Row 3) inferred values from the mathematical model. The mathematical model captures the fact that there are kinetics in the Type II cell and improves on the amplifier’s estimates of the components.

Rk (MΩ) Ck (pF) Rm (MΩ) Cp (pF) Cm (pF) Rs (MΩ) Voff (mV)
component label 100 1000 500 4.7 22 30 0
Patchmaster estimate 91.30 8.80 41.19 33.60 −1.20
fitted parameters 94.20 1062.69 520.70 4.85 36.38 34.87 0.20