Table 1.
DICOM-Confirmable Factor | Protocol | Allowable Change from Baseline (T0) |
---|---|---|
Acquisition Duration | 80–100 seconds | ≤3-second change |
Early Phase Timing | 120–150 seconds post contrast injection | ≤5-second change |
Field of View | 260–360 mm | ≤50-mm change |
In-Plane Resolution | ≤1.4 mm | ≤10% change |
Slice Thickness | ≤2.5 mm | ≤10% change |