Table 3.

PCR results compared to the 87 sequenced samples. Positive was counted, if at least one PCR run (triplicate) was positive. PPV: positive predictive value, NPV: negative predictive value, FNR: false negative rate, FPR: false positive rate, ST: serologic test.

87 Sequenced Samples (Total Number of Test Runs: 261) PCRs
kDNA 18S rRNA TCZ NDO
True positive 174 3 40 180
False positive 51 0 0 0
False negative 21 192 155 15
True negative 15 66 66 66
Sensitivity (%) 89.23 1.53 20.51 92.31
Specificity (%) 22.73 100 100 100
PPV (%) 77.33 100 100 100
NPV (%) 41.67 25.58 29.86 81.48
FNR (%) 10.77 98.46 79.49 7.69
FPR (%) 77.27 0 0 0
Accuracy (%) 72.41 26.44 40.61 94.25