TABLE 1.

Common imaging parameters for the in vivo 2D and 3D T mapping experiments

Parameter 2D mapping 3D mapping
Orientation Axial Sagittal
Phase-encoding directions R/L A/P and R/L
FOV (cm) 22 × 22 24 × 24 × 16
Sampling matrix 256 × 256 256 × 128 × 80
Slice thickness (mm) 10 2
TR/TE (ms) 8.5/3.8 6.8/2.95
Bandwidth (Hz/px) 260 260
Fat saturation Yes Yes
GRAPPA acceleration None Table 2
VPS 32, 64, 128, 256 Table 2
No CSF nulling: Trec (ms) 2500 Table 2
CSF nulling: TIprep/TEprep/TI (ms) 1350/200/940 Table 2
TSLs (ms) 0, 20, 40, 60, 80 Table 2
Spin-lock pulse amplitude (Hz) 350 350
Assumed brain tissue T1 (ms) 1200 1200
Assumed brain tissue T2 (ms) 70 70
Assumed brain tissue T (ms) 75 75