Table 8. Standard deviations of the repeatability and within-lab precision for the assay of NFL concentrations in PBS using the IMR NFL reagent.
Material | Mean of measured NFL concentrations | Standard deviation (Coefficient of variation) | |
---|---|---|---|
Repeatability | Within-lab precision | ||
NFL-PBS sample 1 | 1.00 pg/ml | 0.027 pg/ml (2.67%) | 0.019 pg/ml (1.89%) |
NFL-PBS sample 2 | 100.17 pg/ml | 0.528 pg/ml (0.53%) | 1.313 pg/ml (1.31%) |