Table 2.

Comparison of Pediatric MLung to Current Devices in Clinical Use and Next Generation Devices in the Literature

Comparison Between MLung and Commercial Oxygenators
Device Surface Area (m2) Rated Flow (L/min) Pressure Drop at Rated Flow (mm Hg) Priming Volume (mL) Gas Exchange Efficiency at Rated Flow (mL/min/m2)
O2 CO2
Terumo Capiox RX05 (Terumo Cardiovascular Group, Ann Arbor, Michigan) 0.5 1.5 110 43 150 500*
Maquet Quadrox Neonatal (Maquet, Rastatt, Germany) 0.38 1.5 58 40 197 189
Maquet Quadrox Pediatric (Maquet, Rastatt, Germany) 0.8 2.8 60 (1.5 L/min) 99 175 250
Sorin Dideco D100 (Sorin Group, Saluggia, Italy) 0.22 0.7 175 47 159 145
Novalung ILA (Xenios AG, Heilbronn, Germany) 1.3 Data not available Data not available 175 85 212
Pedipump Lung (University of Maryland, Baltimore, Maryland) 0.3 2.5 150 110 400 343
P-PAL (University of Pittsburgh, Pittsburgh, Pennsylvania) 0.3 Data not available§ Data not available§ 270§ 310§ Data not available
MLung (single inlet); (University of Michigan, Ann Arbor, Michigan) 0.28 2.0 106 47 357 714
Pediatric MLung (University of Michigan, Ann Arbor, Michigan) 0.13 1.0 25 68 ± 7 349 343 ± 90

Data for commercial devices were gathered from manufacturers technical information available via company websites.

*

Capiox RX05 CO2 removal data were estimated from Sun et al.32

Novalung iLA is designed for CO2 removal, and data at rated flow were not available; therefore, the CO2 removal at the maximum flow for which data were available was reported.

CO2 exchange efficiency for the Pedipump Lung was estimated from Wei et al.33

§

Rated flow was not reported for the P-PAL; however, gas exchange efficiency was comparable to the PediPL at a flow rate of 2 L/min.33,34 Priming volume of the P-PAL circuit is reported which includes fiber bundle area, pump area, and conduits.35 Pressure drop of gas exchange region not reported; this device incorporates a pump to generate pressure and is intended for venoarterial use. CO2 removal data not reported.

CO2 exchange efficiency was taken as the average of the highest values achieved during in vivo testing (n = 3).

CO2, carbon dioxide; O2, oxygen.