Table 2:
The number of patients versus the ranges of measurement differences, in addition to the maximum and minimum values of these ranges (in millimeters), (Reprinted from Beshtawi, 2017).[21]
| DP-CRP (%) | DP-CORO (%) | CRP-CORO (%) | |||||
|---|---|---|---|---|---|---|---|
| Difference range | X1 | X2 | MH | MW | Y1 | Y1 | Y1 |
| 0 mm | 2 (7) | 1 (4) | 2 (7) | 1 (4) | 3 (11) | 2 (7) | 1 (4) |
| >0–0.5 mm | 13 (46) | 12 (43) | 17 (61) | 21 (75) | 16 (57) | 5 (18) | 10 (36) |
| 0.6–1 mm | 10 (36) | 5 (18) | 4 (14) | 6 (21) | 4 (14) | 2 (7) | 4 (14) |
| >1 mm | 3 (11) | 10 (36) | 5 (18) | 0 (0) | 5 (18) | 19(68) | 13 (46) |
| Total number of patients | 28 | 28 | 28 | 28 | 28 | 28 | 28 |
| Max. difference | 2.9 mm | 2 mm | 2 mm | 1 mm | 2.6 mm | 5.3 mm | 5 mm |
| Min. difference | 0 mm (No difference) | ||||||
DP-CRP, DP-CORO, CRP-CORO: Distance difference between digital and CRP views, digital panorama and the CBCT coronal view, and CRP and coronal views, respectively. CBCT: Cone-beam computed tomography